Arguments
Returns
Examples
[SOURce[1]]:FREQuency:STOP
Group
Syntax
Related Commands
Arguments
Returns
Examples
AFG2021 Arbitrary Function Generator Programmer Manual
<frequency>::=<NRf>[<units>]
where:
<NRf> is the start frequency.
<units>::=[Hz | kHz | MHz]
<frequency>
SOURce1:FREQuency:STARt 10kHz
sets the sweep start frequency of CH 1 to 10 kHz.
This command sets or queries the stop frequency of sweep for the specified
channel. This command is always used with the [SOURce[1]]:FREQuency:STARt
command. The setting range of stop frequency depends on the waveform selected
for sweep. For more information, refer to the AFG2021 instrument Specifications
and Performance Verification Manual.
Source
[SOURce[1]]:FREQuency:STOP
[SOURce[1]]:FREQuency:STOP
[SOURce[1]]:FREQuency:MODE, [SOURce[1]]:FREQuency:STARt
<frequency>::=<NRf>[<units>]
where:
<NRf> is the stop frequency.
<units>::=[Hz | kHz | MHz]
<frequency>
SOURce1:FREQuency:STOP 100KHz
sets the stop frequency of CH 1 to 100 kHz.
Command Descriptions
{<frequency>|MINimum|MAXimum}
? {MINimum|MAXimum}
2-51