Returns
Examples
[SOURce[1]]:PM[:DEViation]:DCYCle
Group
Syntax
Arguments
Returns
Examples
[SOURce[1]]:PM:INTernal:FREQuency
AFG2021 Arbitrary Function Generator Programmer Manual
<phase>
SOURce1:PHASe:ADJust MAXimum
CH 1 output frequency.
This command sets or queries the peak frequency deviation of PM modulation for
the specified channel.
Source
[SOURce[1]]:PM[:DEViation]:DCYCle
[SOURce[1]]:PM[:DEViation]:DCYCle
<deviation>::=<NR3>[<units>]
where:
<NR3> is the phase deviation.
<units>::=[RAD | DEG]
If <units> are omitted, RAD is specified automatically. The setting ranges are:
RAD: 0 PI to +1 PI, relative to phase value
DEG: 0 to +180, in 0.1 degree steps, relative to phase value
<deviation>
SOURce1:PM:DEViation:DCYCle MAXimum
sets the maximum value for the CH 1 phase deviation.
This command sets or queries the internal modulation frequency of PM
modulation for the specified channel. You can use this command only when the
internal modulation source is selected.
You can set the internal modulation frequency from 2 mHz to 50.00 kHz with
resolution of 1 mHz.
Command Descriptions
sets the maximum value for the phase of
{<deviation>|MINimum|MAXimum}
?
{MINimum|MAXimum}
2-57