Tektronix AFG2021 Programmer's Manual page 82

Arbitrary function generator
Hide thumbs Also See for AFG2021:
Table of Contents

Advertisement

Command Descriptions
Examples
[SOURce[1]]:PULSe:TRANsition[:LEADing]
Group
Syntax
Arguments
Returns
Examples
[SOURce[1]]:PULSe:TRANsition:TRAiling
Group
Syntax
Arguments
2-64
SOURce1:PULSe:PERiod 200ns
sets the CH 1 pulse period to 200 ns.
This command sets or queries the leading edge time of pulse waveform.
Source
[SOURce[1]]:PULSe:TRANsition[:LEADing]
MAXimum}
[SOURce[1]]:PULSe:TRANsition[:LEADing]
<seconds>::=<NRf>[<units>]
where:
<NRf> is the leading edge time of pulse waveform.
<units>::=[ns | μs | ms | s]
<seconds>
SOURce1:PULSe:TRANsition:LEADing 200ns
sets the CH 1 leading edge time to 200 ns.
This command sets or queries the trailing edge time of pulse waveform.
Source
[SOURce[1]]:PULSe:TRANsition:TRAiling
MAXimum}
[SOURce[1]]:PULSe:TRANsition:TRAiling
<seconds>::=<NRf>[<units>]
where:
<NRf> is the trailing edge of pulse waveform.
AFG2021 Arbitrary Function Generator Programmer Manual
{<seconds>|MINimum|
?
{MINimum|MAXimum}
{<seconds>|MINimum|
?
{MINimum|MAXimum}

Advertisement

Table of Contents
loading

Table of Contents