AFG2021 Arbitrary Function Generator Programmer Manual
*OPT?
None
<OPT>[,<OPT>[,<OPT>[,<OPT>]]]
*OPT?
might return 0, which indicates no option is installed in the instrument.
The OUTPut:IMPedance command sets the output load impedance for the
specified channel. The specified value is used for amplitude, offset, and high/low
level settings. You can set the impedance to any value from 1 Ω to 10 kΩ with a
resolution of 1 Ω or 3 digits. The default value is 50 Ω .
The OUTPut:IMPedance? command returns the current load impedance setting
in ohms. If the load impedance is set to INFinity, the query command returns
"9.9E+37".
Output
OUTPut[1]:IMPedance
OUTPut[1]:IMPedance?
<ohms>::=<NR3>[<units>]
where:
<units>::=OHM
INFinity sets the load impedance to >10 kΩ.
MINimum sets the load impedance to 1 Ω.
MAXimum sets the load impedance to 10 kΩ.
<ohms>::=<NR3>
OUTPut1:IMPedance MAXimum
sets the CH 1 load impedance to 10 kΩ .
{<ohms>|INFinity|MINimum|MAXimum}
{MINimum|MAXimum}