Group
Syntax
Arguments
Returns
Examples
[SOURce[1]]:PULSe:DCYCle
Group
Syntax
Related Commands
AFG2021 Arbitrary Function Generator Programmer Manual
This command is available when Run Mode is set to Continuous, Burst, or Sweep.
You can set or query whether to add the internal noise to the output signal using
the [SOURce[1]]:COMBine:FEED command.
Source
SOURce<3>:POWer[:LEVel][:IMMediate][:AMPLitude]
{<percent>|MINimum|MAXimum}
SOURce<3>:POWer[:LEVel][:IMMediate][:AMPLitude]
{MINimum|MAXimum}
<percent>::=<NR2>[<units>]
where:
<NR2> is the noise level.
<units>::=PCT
<percent>
SOURce3:POWer:LEVel:IMMediate:AMPLitude 50PCT
sets the internal noise level that is added to the output signal to 50%.
This command sets or queries the duty cycle of the pulse waveform for the
specified channel. The setting range is 0.001% to 99.999% in increments of 0.001.
The instrument will hold the settings of leading edge and trailing edge when
the duty cycle is varied.
Refer to the [SOURce[1]]:PULSe:WIDTh command for the setting range.
Source
[SOURce[1]]:PULSe:DCYCle
[SOURce[1]]:PULSe:DCYCle
[SOURce[1]]:PULSe:WIDTh
{<percent>|MINimum|MAXimum}
?
{MINimum|MAXimum}
Command Descriptions
?
2-61