[SOURce[1]]:FREQuency[:CW|:FIXed]
Group
Syntax
Arguments
2-48
Source
[SOURce[1]]:FREQuency:CENTer
[SOURce[1]]:FREQuency:CENTer
[SOURce[1]]:FREQuency:SPAN,
<frequency>::=<NRf>[<units>]
where:
<NRf> is the center frequency.
<units>::=[Hz | kHz | MHz]
<frequency>
SOURce1:FREQuency:CENTer 550kHz
sets the CH 1 center frequency to 550 kHz.
This command sets or queries the frequency of output waveform for the specified
channel. This command is available when the Run Mode is set to other than
Sweep.
The setting range of output frequency depends on the type of output waveform. If
you change the type of output waveform, it might change the output frequency
because changing waveform types impacts on the setting range of output
frequency. The resolution is 1 μHz or 12 digits. For more information, refer to the
AFG2021 instrument Specifications and Performance Verification Manual.
Source
[SOURce[1]]:FREQuency[:CW|:FIXed]
[SOURce[1]]:FREQuency[:CW|:FIXed]
<frequency>::=<NRf>[<units>]
where:
<NRf> is the output frequency.
AFG2021 Arbitrary Function Generator Programmer Manual
{<frequency>|MINimum|MAXimum}
?
{MINimum|MAXimum}
[SOURce[1]]:FREQuency:MODE
{<frequency>|MINimum|MAXimum}
?
{MINimum|MAXimum}