Ultima X Series Unit Calibration End Display; Span Gas Connection (Ultima Unit Shown) - Hach sc200 User Manual

Hide thumbs Also See for sc200:
Table of Contents

Advertisement

12.After the 30 second countdown:
• The display alternates between "CAL" and a value.
(for example: 60 PPM for 0 to 100 ppm carbon monoxide).
This value is the actual reading of the gas concentration the
sensor is detecting. The engineering units (PPM, %, or %LEL)
are predetermined by the type of sensor installed and are
not changeable.
• Once the gas value is stable, the alternating display stops. If
the calibration is successful, the display will show END for
approximately two seconds. (FIGURE 2-7).
Figure 2-7. Ultima X Series Unit Calibration End Display
No user adjustments are necessary.
The display shows the span gas value while the span gas is
flowing to the unit. (For example, it may read 60 PPM or 25%
or 60% LEL).
13.Turn OFF gas flow by turning the knob on the flow controller.
Figure 2-8. Span Gas Connection (Ultima unit shown)
If the calibration output signal is enabled during calibration, it
will be held at the lockout value for two additional minutes after
END is displayed.
2-15