Test And Maintenance Menu - Hach sc200 User Manual

Hide thumbs Also See for sc200:
Table of Contents

Advertisement

Problem
Sensor not recognized
ote
a ple of possi le display
essage
Device Missing error message
appears

Test and Maintenance menu

1. From the Settings Menu, select Test/Maint and push ENTER.
Option
Description
Scan devices
Performs a scan for active and missing devices.
44
nglis
Resolution
If the sensor is an analog sensor and a
corresponding module is installed in the
controller, refer to the instructions supplied
with the Network or Sensor Module.
Make sure the digital connector wiring
harness is seated on the inside of the door
assembly and that the wiring harness is not
damaged.
If the digital sensor is connected to the
controller with a digital termination box, user
supplied junction box, digital extension
cables, or a user-supplied extension cable,
connect the sensor directly to the controller
and perform a device scan. If the controller
recognizes the sensor, check that all the
wiring in the junction boxes or extension
cables is correct.
Make sure that only two sensors are installed
in the controller. Although two analog module
ports are available, if a digital sensor and two
analog modules are installed, only two of the
three devices will be seen by the controller.
Contact Technical Support
Perform a Device Scan from the
Test/Maintenance menu.
Power cycle the controller
Option
Description
Output cal
Lets the user calibrate the 4–20 mA outputs with a
250 ohm resistor in series to the mA output terminals.
Output 1
The settings for each output are adjusted until the
correct value (4 mA or 20 mA) is supplied.
Output 2
Calibrate 4 mA output (Min: 0 Max: 25000)
Calibrate 20 mA output (Min: 35000 Max: 65533)
Hold outputs
Sets the value the controller sends to an external
system for a defined period of time. After the time
period, the instrument goes back to reporting real time
values.
Set activation—Launch or release
Set Outmode—Hold Outputs (default) or Transfer
Outputs
Set Channels—All (default) or select from hardware
list
Test output
Lets the user select a mA value that is sent by the
controller for verification.
Output 1
Min: 0 mA (default +04.00)
Output 2
Max: 25.00 mA
Status
View status of all modules, sensors and relays.
Test relay A, B,
Energize or De-energize the selected relay
C, D
Overfeed reset
Resets the Overfeed Timer.
Reset default
Resets the controller configuration settings to the
config
default values (language, date and time, relay function
and data output function).
Restart sc200
Performs a controller restart