6. To disable the calibration signal, aim the controller at the sensor
and press the 0 button.
The command is sent immediately
The Ultima Gas Monitor flashes Sig OFF
The Ultima X Series Gas Monitor displays: CAL SIG OFF.
Procedure 9. Setting the Number of Days
Between Ultima Gas Monitor Auto-calibration Periods
With the use of the Ultima Gas Monitor auto-calibration module, the
Ultima Gas Monitor is capable of automatically applying zero and span
gas to itself. This provides a complete calibration of the sensor without
operator intervention. If the Ultima Gas Monitor auto-calibration module
is connected, the Ultima Gas Monitor must be programmed to calibrate
itself between 1- to 128-day intervals.
1. Press the SEND button.
The display prompts: Send?
2. Press the CAL button.
The display prompts: Sel Cal Action +.
3. Press either the + or - button and scroll to the DaysPerAutoCal
display message.
4. Press the ENTER button.
The display prompts: CalTerm ### dy
5. Enter the three-digit period desired (from 1 day to 128 days).
(Leading zeros are required.)
6. Aim the controller at the sensor and press the ENTER button.
The Ultima Gas Monitor shows the number of days between
auto-calibrations for five seconds
The Ultima Gas Monitor is now programmed to auto-calibrate
at the desired interval.

Setting Time that Ultima Gas Monitor Auto-calibration is to begin:

1. Press the SEND button.
The display prompts SEND?
2.
Press the CAL button.
The display prompts: Sel Cal Action +.
3. Press either the + or - button and scroll to the StartHr of Cal
display message.
3-13