Section 3
Installation
DANGER
Only qualified personnel should conduct the tasks described in this section of the
manual.
DANGER
Seul un technicien qualifié peut effectuer les tâches d'installation décrites dans
cette section du manuel.

3.1 Connecting/Wiring the Sensor to the sc100 Controller

DANGER
The sc100 and certain versions of the sensor are suitable for use in Class 1,
Division 2, Groups A, B, C, D Hazardous Locations . See Control Drawing 58600-78
in the sc100 Controller Manual, Cat. No. 58600-18 for acceptable sensor versions
and installation requirements.
DANGER
Le sc100 et certaines versions du capteur peuvent être utilisés dans des endroits
dangereux de la Classe 1, Division 2, Groupes A, B, C, D. Reportez-vous au schéma
de contrôle 58600-78 du Manuel du contrôleur sc100, Réf. 58600-18 pour connaître
les versions des capteurs admises et les conditions d'installation.

3.1.1 Connecting the sc Sensor to a sc100 Controller in a Non-hazardous Location

3.1.1.1 Attaching a sc Sensor with a Quick-connect Fitting

Important Note: The standard quick-connect fitting is NOT suitable for Class 1, Division 2
Hazardous Location installations without the connector lock installed, see
page 13
The sensor has a keyed quick-connect fitting for easy attachment to the controller
(Figure
removed. Extension cables may be purchased to extend the sensor cable length. If the
total cable length exceeds 100 m (300 ft), a termination box must be installed. See
Replacement Parts and Accessories on page
for more information.
5). Retain the connector cap to seal the connector opening when the sensor is
41.
section 3.1.2 on
11