Note: The probe errors
and the reference timer
time-out require the cause
of the fault to be corrected
before they will clear.
2.3

Calibration

SECTION 9 TROUBLESHOOTING for additional information on the system warnings.
To display or clear the WARNINGS, press
Press the
key to scroll to:
UP ARROW
Press
.
ENTER
Press
. The system displays the Warning Codes shown in Table 3. Additional details
ENTER
about the warning can be found in Section 9.1.2 Error Codes and Section 9.1.3 System
Warnings. The warning codes appear momentarily followed by CLeaR WaRNing.
Press
to CLeaR WaRNings on the display.
ENTER
Use this menu to calibrate the pH (or ORP) electrode, review the pH calibration, calibrate
the temperature sensor, and adjust the mV offset. The calibration menu is entered by
pressing the CAL key. The first display to appear is STD 1, followed by STD 2 (after STD
1 is completed), GRAB SMP, REVIEW, TEMP, and MV ZERO. The user may then scroll
up or down through the first level of the menu to make adjustments to each calibration.
(Note: If the keyboard is locked, only the REVIEW sub menu may be entered.)
to call up the setup menu:
SETUP
Table 3 Warning Codes
CODE
MESSAGE
E00
EE FAIL
E01
REF TIME
E02
SOFT SRT
E03
COLD SRT
E04
PROBE ER
E05
PROBE ER
E06
PROBE ER
E07
PROBE ER
E08
MARG CAL
E09
MARG CAL
E10
MARG CAL
E11
MARG CAL
E12
PWR FAIL
37
CAUSE
EEPROM failure
reference timer time-out
software restart
cold restart
pH probe open
pH probe shorted
temperature probe open
temperature probe shorted
marginal pH offset
marginal pH gain
marginal temperature offset
marginal temperature gain
power failure