RayTek MI3 Operating Instructions Manual page 91

Hide thumbs Also See for MI3:
Table of Contents

Advertisement

Start
address
100
110
120
130
410
420
430
5<k>0
5<k>1
5<k>2
5<k>3
5<k>5
5<k>7
MI3
info@Raytek-Direct.ca
Size
Modbus Access Data
[bits]
Type
8
discretes input
bit field Get connected heads
8
discretes input
bit field Get registered heads
16
holding register short
16
holding register short
32
input register
float
32
input register
float
16
input register
short
16
holding register short
16
holding register short
16
holding register short
32
holding register float
32
holding register float
32
holding register float
Content
Values
bit 0: head 1 ..
bit 7: head 8
bit high: head connected
bit low: head disconnected
bit 0: head 1 ..
bit 7: head 8
bit high: head registered
bit low: head not registered
Laser control (only with
0: off, 1: on
laser)
Relay alarm output
0: off
control
1: on
2: norm. open
3: norm. closed
analog input 1 value
0 .. 5 V
analog input 2 value
0 .. 5 V
Trigger
0: off
1: on
analog output k mode
5: TCJ
6: TCK
7: TCR
8: TCS
9: 0...5 V
10: 0...10 V
99: disable (tristate)
analog output k source head number or 0:
fixed value from 5<k>3
analog output k source
1: internal temp. of 5<k>1
parameter
2: object temp. of 5<k>1
analog output k fixed
value within range set in address 5<k>0 -
temp. value
analog output k bottom
device bottom temp. .. device top temp. L<k>O
temp. value
analog output k top
device bottom temp. .. device top temp. H<k>O
temp. value
Rev. F1 05/2013
Modbus
MI3[M]
command
HC
HCR
XL
KB
TV1I
TV2I
XT
XO<k>O
O<k>O
O<k>O
91
1.800.868.7495

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MI3 and is the answer not in the manual?

Questions and answers

Table of Contents