ASCII Programming
Description
Presel. Emissivity
Pointer
Emissivity Source
Presel. Emissivity Value EV
Valley hold time
1
Flicker Filter
Average time
2
Head Address
Connected heads
Registered heads
Head Status Code
Head identification
Top temperature value
of output 1
Top temperature value
of output 2
Laser Control
4
Head serial number
Head special
Head Firmware Revision HV
setting average / peak / valley / advanced hold cancels all other hold modes
1
setting average / peak / valley / advanced hold cancels all other hold modes
2
Head Status Codes (read only):
3
Temperature unit
Object temperature out of range
Ambient temperature out of range
Parameter error for a command
Register write error
Self-test error
{reserved}
Ambient temperature compensation
for MI3100 heads only
4
110
info@Raytek-Direct.ca
Char
Format
P B S N Legal Values
EP
nEP
integer
ES
nES
√
integer
nEV
√
float
F
nF
√ √ √ √ 0.0 - 998.9 s
float
FF
nFF
integer
G
nG
√ √ √ √ 0 – 999.0 s
float
HA
nHA
√
HC
string
√
HCR
string
√
HEC
nHEC
√
16 bit hex
HI
?nHI
√
string
H1O
float
√
(H)
H2O
float
√
HL
nHL
√
integer
HN
?nHN
√
integer
HS
?nHS string
√
?nHV
√
BIT0 // 0 = °C, 1 = °F
BIT1 // 1 = out of range
BIT2 // 1 = out of range
BIT3 // 1 = error
BIT4 // 1 = error
BIT5 // 1 = error
BIT6
BIT7 // 0 = off, 1 = on
Rev. F1 05/2013
0 - 7
I = Emissivity from Internal
√
(by command)
E = Emissivity from
External analog input
(0V - 5V)
D = digital selected FTC1-3
0.1 - 1.1
√
(999 = ∞)
0...32768
1 – 9
e.g. !HC – no heads
!HC1 2 3 7 8
e.g. !HCR – no heads
√
!HCR1 2 3 7 8
HCR=0 –> new registration
hex value of Status Code
e.g. !7HIMIXLTS22
(√
)
√ √ H1O= –40.0
or H= –40.0
√ √ H2O= –40.0
√ √ 0=off, 1=on, 2=flash,
3=external
e.g. !nHN98123
√
e.g. !99HSRAY
√
e.g. 1.01
Factory default Head /
Box
0
H
I
H
H
0.0 s
H
0 (LT, G5)
H
3000 (1M, 2M)
0.0 s
H
B
B
B
H
3
set at production H
500°C
B
500°C
B
0
H
set at production H
Set at
H
production
Set in FW
H
MI3
1.800.868.7495
Need help?
Do you have a question about the MI3 and is the answer not in the manual?
Questions and answers