Printing RTD Data

4-12
Use the RTDs submenu to send the numeric point value of a
connected RTD device to the RS232 port for display on a terminal or
RS232 printer. By viewing the point value of the meter at a given
temperature, you can calculate the points at the Set Temperature, and
the slope of the RTD input in points/degree. (See Chapter 1,
Calibration, of the Service Manual for more details.)
1. To access the Print Menu, press the
Note: The Print Menu is not protected by the security feature and a
password is never required to access this menu.
2. Press the [←] or [→] key to display the option bar shown, and
then press
to select the
[F3]
3. At the Dump RTD Data prompt, press
menu without dumping data, or
RS232 port.
After the selection has been made, the Model GF868 returns to the
standard data display screen and normal measurement taking. If you
have selected YES at the DUMP RTD DATA prompt, the meter
continues to send data until you reenter the PRINT menu and select
the STOP option.
[PRNT]
key.
submenu.
RTDs
[F1]
to leave the
to send the RTD data to the
[F2]
PRINT
Printing Data