Setting Up Rtd Inputs; Entering The Zero Cutoff - GE Panametrics GF868 Programming Manual

Flare gas flow meter
Table of Contents

Advertisement

March 2003

Setting up RTD Inputs

Entering the Zero Cutoff

1-20
Option cards with RTD inputs have a temperature range of –100° to
350°C. Complete the following steps to set up the two RTD inputs of
an option card installed in Slot x:
1. Press
[F1]
to set up RTD input A or
Note: The setup of RTD input A is used as an example. Identical
procedures would be used to set up RTD input B.
2. Enter a label of up to eight characters and press
3. At the Input Measurement prompt, press
RTD input A and return to the previous prompt, or press
TEMP to enable RTD input A.
4. Enter the Zero Value for the low end of the chosen input range and
press
.
[ENT]
5. Enter the Full Scale Value for the high end of the chosen input
range and press
[ENT]
6. Press
[F1]
-
[F2]
to set up another input or press
another slot for set up.
For instructions on setting up additional inputs and/or outputs
proceed to the appropriate section. Otherwise, press
to the main
I/O
menu prompt.
Near zero flow, the Model GF868's readings may fluctuate due to
small offsets caused by thermal drift or similar factors. To force a
zero reading when there is minimal flow, enter a zero cutoff value as
described below:
1. At the main I/O menu prompt, press
2. Enter the desired Zero Cutoff value and press the
value of 0.1 ft/s (0.03 m/s) is recommended, but values from 0–1
ft/s (0–0.3 m/s) are acceptable. The menu returns to the main I/O
prompt.
[F2]
to set up RTD input B.
= OFF to disable
[F1]
.
[F3] = ZERO.
Programming Site Data
.
[ENT]
=
[F2]
[EXIT]
to select
[EXIT]
to return
key. A
[ENT]