Setting Up Analog Outputs - GE Panametrics GF868 Programming Manual

Flare gas flow meter
Table of Contents

Advertisement

March 2003
Setting Up Analog
Outputs
1-16
The Model GF868 has two built-in analog outputs, which are
assigned to
. Also, a variety of option cards may be installed in
Slot 0
the six expansion slots. See Chapter 1, Installation, of the Startup
Guide for a complete description of the available option cards.
This menu option is used to set up and/or scale the analog inputs and
outputs. To accomplish this, complete the following steps:
I/O
I/O
last selection appears here
SLOT0 SLOT1 SLOT2 SLOT3
Note: If the number of a slot with an installed option card does not
appear, the card may not have been initialized or it may be
defective. Call the factory for assistance.
Complete the following steps to set up the
1. Use the
[F1]-[F4]
keys to set up outputs A, B, C or D, respectively.
Note: The set up of output A is used as an example. Identical
procedures would be used to set up the other outputs.
2. Press
to disable output A and return to the previous
[F1] = OFF
prompt, or press
[F2] = 0-20 m or [F3]
desired range for output A.
3. Use the [←] and [→] and
Output Measurement parameter, as listed in Table 1-4 below.
Table 1-4: Output Measurement Options
Option Bar Choice
=
[F1]
VEL
=
[F2]
VOLUM
[F3]
=
+TOTL
=
[F4]
-TOTL
→]
=
[
+ [F1]
MDOT
→]
[
+ [F2]
=
+MASS
→]
[
+ [F3]
=
-MASS
→]
=
[
+ [F4]
DIAG
Use the [←] and [→] and the
[F1]-[F4]
keys to select the
desired slot number. (Only those
slots which contain an option
card will appear on the option
bar.)
analog outputs:
Slot x
= 4-20 m to specify the
[F1]-[F4]
keys to specify the desired
Description
Flow Velocity
Volumetric Flow
Forward Totalized Volume Flow
Reverse Totalized Volume Flow
Mass Flow
Forward Totalized Mass Flow
Reverse Totalized Mass Flow
Diagnostic
Programming Site Data