Option Card Alarms - GE Panametrics GF868 Programming Manual

Flare gas flow meter
Table of Contents

Advertisement

Setting Up Analog
Outputs (cont.)

Option Card Alarms

Programming Site Data
The
DIAG
option provides access to a wide variety of diagnostics
parameters. See Chapter 3, Diagnostics, of the Service Manual for a
complete description of these options.
4. Enter the Zero value for the low end of the chosen output range.
Press
[ENT]
.
5. Enter the Full value for the high end of the chosen output range.
Press
[ENT]
.
6. Use the
[F1]
-
[F4]
keys to set up another output or press
select another slot for setup.
For instructions on setting up additional inputs and/or outputs
proceed to the appropriate section. Otherwise, press
to the main
I/O
menu prompt.
1. Use the
[F1]-[F3]
keys to set up alarm relays A, B, or C,
respectively.
Note: The set up of alarm A is used as an example. Identical
procedures would be used to set up the other alarms.
2. Press
= OFF to disable Alarm A and return to the previous
[F1]
prompt, or press
[F2] = HIGH, [F3] = LOW or [F4]
specify the type for Alarm A.
3. Press [F1] = NO for standard operation or [F2] = YES for
Failsafe operation.
4. Do one of the following:
If you selected FAULT, proceed to step 5.
If you selected HIGH or LOW, complete the steps below.
a. Use the [←] and [→] and
Output Measurement parameter, as shown in Table 1-7 on the
previous page.
b. Enter a value for the trigger point of the alarm and press
5. To select the Fault Type that will trigger the alarm, press
FLOW
,
[F2] = OTHER
6. Press
-
to set up another alarm or press
[F1]
[F3]
another slot for set up.
For instructions on setting up additional inputs and/or outputs
proceed to the appropriate section. Otherwise, press
to the main
I/O
menu prompt.
keys to specify the desired
[F1]-[F4]
(non-flow), or
[F3]= ALL
March 2003
[EXIT]
to
to return
[EXIT]
= FAULT to
.
[ENT]
[F1] =
.
to select
[EXIT]
to return
[EXIT]
1-17