Introduction
This section specifies the operating data of the components in ÄKTA avant. For general
data for the system see Section 9.1 System specifications, on page 422.
System pumps
Parameter
Pump type
Flow rate range (normal range)
Flow rate range (column packing flow)
Pressure range
Viscosity range
Flow rate
specifications
424
Data
Piston pump, metering type
ÄKTA avant 25: 0.001 to 25 ml/min
ÄKTA avant 150: 0.01 to 150 ml/min
ÄKTA avant 25: 0.01 to 50 ml/min
ÄKTA avant 150: 0.01 to 300 ml/min
ÄKTA avant 25: 0 to 20 MPa (0 to 200 bar)
ÄKTA avant 150: 0 to 5 MPa (0 to 50 bar)
ÄKTA avant 25: 0.35 to 10 cP
ÄKTA avant 150: 0.35 to 5 cP
ÄKTA avant 25:
Accuracy: ± 1.2%
Precision: RSD < 0.5%
(Conditions: 0.25 to 25 ml/min, < 3 MPa, 0.8 to 2 cP)
ÄKTA avant 150:
Accuracy: ± 1.5%
RSD < 0.5%
(Conditions: 1.0 to 150 ml/min, < 3 MPa, 0.8 to 2 cP)