8 Troubleshooting

8.7 Troubleshooting: Other components

8.7
Troubleshooting: Other components
In this section
General hardware: All modules
Problem
Siphoning occurs
through the system in
column and flow restric-
tor by-pass
Modules cannot be
found by the instru-
ment
404
General Hardware: All modules
Mixer
Superloop
Cabinet
Power and ICU
External air sensors
I/O-box E9
Instrument display
Possible cause and action
The internal pump flow restrictor needs to be replaced,
see Section 7.4.2 Check the function of the pump flow
restrictors, on page 281.
The cable between a module and the ICU is not
connected
Remove the module and make sure that the cable is
connected.
Two similar modules have been added to the instru-
ment, for example two Inlet valve A
Remove one of the modules with the same Node ID.
The Node ID for one or more of the modules is incor-
rect, for example an Inlet valve A2 has the same
Node ID as Inlet valve A. The instrument then con-
siders them to have the same identity.
Remove Inlet valve A2 and change the Node ID accord-
ing to Section 9.15 Check and change the Node ID of
a module, on page 515.