9 Reference information
9.7 System settings

9.7.1 System settings - UV

9.7.1
System settings - UV
The following table describes the UV related system settings available for ÄKTA avant.
Instruction name
Alarm UV1
Alarm UV 2nd
Wavelength
Noise reduction UV
Noise reduction UV 2nd
460
Description
Enables or disables the alarm for the UV 1 signal from
UV monitor U9-M. When enabled, it sets the alarm limits
for the UV 1 signal. When the signal falls outside the set
limits, an alarm is issued and the method will be paused.
Note:
It is not possible to set an alarm signal for the UV 2 or
UV 3 signals from UV monitor U9-M.
Enables or disables the alarm for the UV signal from U9-
L, 2nd. When enabled, it sets the alarm limits for the UV
signal from UV monitor U9-L, 2nd. When the signal falls
outside the set limits, an alarm will be triggered and the
method will be paused.
Sets the wavelengths for UV monitor U9-M. The corre-
sponding curves in the chromatogram are marked with
the set wavelengths.
Note:
The instruction is available when UV monitor U9-M
(variable) is selected in the component list. For best
performance, do not use more wavelengths than
necessary for the application.
At low wavelengths, the eluent can have significant
absorbance of its own.
Filters the noise in the UV signal from U9-M or U9-L
(depending on the configuration). A column-specific
averaging time is set automatically when a column is
defined in a method run and Averaging time is set as
a variable.
Filters the noise in the UV signal from U9-L, 2nd. A col-
umn-specific averaging time is set automatically when
a column is defined in a method run and Averaging
time is set as a variable.
ÄKTA avant User Manual 29035184 AE