7.8.11
Replace pump pistons, piston seals, O-rings and rinse
membranes
Introduction
Pump pistons, piston seals, O-rings and rinse membranes are pump parts that can be
replaced by the user. Replacing these parts require special attention not to damage the
pumps. See ÄKTA avant Operating Instructions for instructions on replacement procedures.
ÄKTA avant User Manual 29035184 AE

7.8.11 Replace pump pistons, piston seals, O-rings and rinse membranes

NOTICE
Do not disassemble the pump head unless there is good reason
to believe that there is an internal leakage. A sign of leakage
is increasing pump rinsing solution volume. Always make sure
that sufficient spare components are available before attempt-
ing to replace a spare part.
Replacing spare parts. Read the instructions carefully. For
example, some individual parts of the pump head can be as-
sembled incorrectly. Check the orientation of each part before
continuing with the next instruction.