7 Maintenance
7.6 Maintenance when required

7.6.12 Clean the Conductivity flow cell

7.6.12

Clean the Conductivity flow cell

Maintenance interval
Clean the Conductivity flow cell when required.
Required material
The following materials are required:
Instruction
Follow the instruction to clean the flow cell of the Conductivity monitor. The same proce-
dure also applies to the second Conductivity monitor.
Step
1
2
3
318
Luer connector
Waste container
Syringe, 25 ml
1 M NaOH
Distilled water
Action
Disconnect the Fingertight connector and the piece of tubing from the top
of the Conductivity monitor, and attach a Luer connector.
Disconnect the piece of tubing from the bottom of the Conductivity monitor,
and connect a piece of waste tubing to the Conductivity monitor. Place the
waste tubing in a waste container.
Disconnect the Conductivity monitor from the rails, or slide the Conductivity
monitor to the right on the rails.