Sample Application Using A Sample Loop - GE AKTA avant User Manual

Hide thumbs Also See for AKTA avant:
Table of Contents

Advertisement

5.7.3

Sample application using a sample loop

Introduction
A sample loop is recommended for injection of smaller sample volumes onto the column.
A sample loop can be connected to either the Injection valve or the Loop valve. When
using the Loop valve, up to five loops can be connected simultaneously.
Note:
How to fill a sample loop
Follow the instructions to fill the sample loop with sample.
Step
1
2
3
ÄKTA avant User Manual 29035184 AE
Sample loop is called capillary loop in UNICORN.
Action
Connect a suitable sample loop to Injection valve ports LoopF (fill) and LoopE
(empty).
Fill a syringe with sample.
Connect the syringe to the Injection valve port Syr.
5.7 Sample application
5.7.3 Sample application using a sample loop
5 Operation