9.7 System settings

9.7.10 System settings - Data collection

9.7.10
System settings - Data collection
The following table describes the data collection related system settings available for
ÄKTA avant.
Instruction name
Name of given curve. E.g.,
UV1, UV2, Cond
474
Description
The Maximum number of data points parameter deter-
mine the maximum number of data points collected for
a given curve. Data reduction occurs if the maximum
number of data points is exceeded. To avoid data reduc-
tion, set the maximum number of data points to be
collected to 180000 or insert a New Chromatogram
instruction in the method.
Note:
The default setting is 54000 data points, which corre-
sponds to 1.5 h for a signal of 10 Hz.