System pumps
Parameter
Pump type
Flow rate range
Pressure range
Viscosity range
Flow rate
specifications
ÄKTA pure User Manual 29119969 AB
Data
Piston pump, metering type
ÄKTA pure 25:
0.001 to 25 ml/min
(up to 50 ml/min during column packing)
ÄKTA pure 150:
0.01 to 150 ml/min
(up to 300 ml/min during column packing)
ÄKTA pure 25:
0 to 20 MPa (2900 psi)
ÄKTA pure 150:
0 to 5 MPa (725 psi)
ÄKTA pure 25:
0.35 to 10 cP (5 cP above 12.5 ml/min)
ÄKTA pure 150:
0.35 to 5 cP
ÄKTA pure 25:
• Accuracy: ± 1.2%
• Precision: RSD < 0.5%
(Conditions: 0.25 to 25 ml/min, < 3 MPa, 0.8 to 2 cP)
ÄKTA pure 150:
• Accuracy: ± 1.5%
• Precision: RSD < 0.5%
(Conditions: 1.0 to 150 ml/min, < 3 MPa, 0.8 to 2 cP)