Specifications; Ambient Environment; Compressed Air - GE AKTAprocess Operating Instructions Manual

Hide thumbs Also See for AKTAprocess:
Table of Contents

Advertisement

8 Reference information

8.1 Specifications

8.1

Specifications

Introduction
This section contains technical data pertaining to the ÄKTAprocess and its components.
For complete technical data, refer to the GS in the product documentation package.
General technical data
The following table provides general technical data for the complete system.
Property
Process operating
limits
Electric power

Compressed air

Ingress protection EN
60529
Noise level
Floor inclination
Altitude

Ambient environment

164
Value
Inlet feed pressure: 0 to 0.2 bar (g)
Outlet back pressure: 0 to 1 bar (g)
Flammable liquids: Not allowed
Operation in EX-rated environment: Not allowed
Refer to Cabinet type specific data, on page 165
Pressure 5.5 to 7 bar
Airflow min. 50 Nl/min during normal operation
System: IP55
Cabinet: IP56, NEMA 4X
≤ 72 dBA
Horizontal +/- 1º
Up to 2000 m
Temperature: 2ºC to 26ºC
Air humidity: 20% to 95%, non-condensing
ÄKTAprocess Operating Instructions 29-0252-49 AA