4.10 Second Conductivity Monitor - GE AKTA avant User Manual

Hide thumbs Also See for AKTA avant:
Table of Contents

Advertisement

4.10 Second Conductivity monitor

Introduction
It is possible to use two conductivity monitors. The second Conductivity monitor could
for example be placed before the column for extended gradient control or on a sample
inlet line for control of ionic strength in the sample.
Location
The recommended position of the second Conductivity monitor is dependent on column
size or where the Sample inlet valve is placed. The nearest empty module position should
be used. It is possible to install the module in an Extension box. See Section 2.6.3 Extension
box, on page 53.
Connect tubing
When connecting a second Conductivity monitor to ÄKTA avant use the tubing dimensions
that are most suitable for the chosen location and application.
System properties
Follow the instruction to update the system properties.
Step
1
2
3
ÄKTA avant User Manual 29035184 AE
NOTICE
UV and conductivity flow cells on the high pressure side. When
placing UV and/or conductivity flow cells on the high pressure side
of the column, the UV flow cell has a maximum pressure limit of 2
MPa (20 bar) and the conductivity flow cell has a maximum pressure
limit of 5 MPa (50 bar).
Action
Open the system properties Edit dialog.
In the Component types list, click Monitors and sensors.
Select the Conductivity monitor 2nd (C9) check box in the Component se-
lection list. Then click OK.
4 Optional instrument modules

4.10 Second Conductivity monitor

137