7.5.1

Clean the UV flow cell

Maintenance interval
Clean the UV flow cell every six months, or when required.
Required material
The following materials are required:
Instruction
Follow the instructions below to clean the UV flow cell of UV Monitor U9-M or UV Monitor
U9-L. The illustrations show UV Monitor U9-M, UV Monitor U9-L is cleaned in the same
way. The UV flow cell can be mounted or not mounted on the instrument during the
cleaning procedure.
ÄKTA avant User Manual 29035184 AE
NOTICE
Keep UV flow cell clean. Do not allow solutions containing dissolved
salts, proteins or other solid solutes to dry out in the flow cell. Do
not allow particles to enter the flow cell, as damage to the flow cell
may occur.
Luer connector
Waste container
Syringe, 25 to 30 ml
10% surfactant detergent solution (e.g., Decon™ 90, Deconex 11, or RBS 25)
Distilled water
7 Maintenance
7.5 Semiannual maintenance
7.5.1 Clean the UV flow cell
283