8.4.3.5
7.4.3.5
ROUTe:SCAN?
Returns a comma delimited list of the enabled channels.
ROUT:SCAN?
(@2,3,4)
The list of channels is enclosed in parenthesis and preceded with the @ symbol.
7.4.3.6
ROUTe:SCAN <chn list>
Sets the channels for scanning.
ROUT:SCAN 2,3
The <chn list> parameter is a comma delimited list of the channels (1 to 4) to enable.
Channels that are omitted from this list are disabled. If the <chn list> parameter is
omitted, all channels are disabled. The *RST command sets channel 1 ON, and all
other channels OFF.
7.4.3.7
ROUTe:SCAN:MODE? [MIN|MAX|DEF]
Returns the current scan mode setting.
ROUT:SCAN:MODE? MIN
0
If 0 is returned, the scan mode is simultaneous. If a 1 is returned the scan mode is
scan. Specifying the MIN parameter returns a value of 0. Specifying the MAX param-
eter returns a value of 1. Specifying the DEF parameter returns a value of 1.
7.4.3.8
ROUTe:SCAN:MODE <num>|MIN|MAX|DEF
Sets the scan mode setting.
ROUT:SCAN:MODE 1
The <num> specifies the scan mode (1 for scan, 0 for simultaneous). Specifying the
MIN, MAX, or DEF parameter sets the mode to the minimum, maximum, or default,
respectively. The *RST command sets scan mode to scan.
7.4.4

Probe Commands

The probe commands are used for querying, setting, and copying the probe conver-
sion type, conversion parameters, and serial number and for testing the temperature