DATA WILL BE LOGGED NO FASTER THAN
THE MEASURE PERIOD (X SEC ).
Press ENTER to continue...
The X in the message is the current measurement period. After the Enter button is
pressed, the measurements are logged at the current measure period.
The COUNT parameter is for selecting the total number of readings to log. The default
is 8,160, which is the maximum number of readings that can be logged. If the count is
set to 10 and two channels are displayed on the screen at the time logging starts, five
readings from each channel are logged for a total of 10 readings. The auto log auto-
matically stops after the number of readings selected is reached or when the user stops
the log manually.
Press the LR buttons to change a setting and the UD buttons to scroll between the
parameters. Press Enter button to save the new setting. Press and hold the Exit button
to cancel to the main display or press the EXIT button to cancel and to move to the
next parameter.
6.4.2.2
Start Stop
The START STOP function (Figure 34 on this page) starts and stops auto log sessions.
When this function is selected, the user is prompted to either start or stop an auto log
session by pressing the Enter button. If an auto log session is not currently running,
the user is prompted to START RECORDING DATA. If an auto log session is run-
ning, the user is prompted to STOP RECORDING DATA.
Figure 34
Start Stop
Note: The log is not started or stopped until the Enter button is pressed to con-
firm the user action.
24.159C
1
24.142C
3
START RECORDING DATA.
Press ENTER to continue...