1529 Chub-E4 Thermometer Readout
Commands
Specifying the MIN, MAX, or DEF parameter sets the scan sequence timer to the
minimum, maximum, or default value respectively. The *RST command sets the scan
sequence timer to 1 second.
7.4.3

Channel Commands

The channel commands are used for querying and setting the current channel, the
channels to scan, the type of scanning, and the state of the channel scanning. The
channel commands are not password protected.
7.4.3.1
ROUTe:CLOSe? <chn>
Returns the state of the specified channel.
ROUT:CLOS? 1
0
The <chn> parameter specifies the channel (1 to 4). If the specified channel is enabled,
this query returns 1. If the specified channel is disabled, this query returns 0.
7.4.3.2
ROUTe:CLOSe <chn>
Enables the specified channel.
ROUT:CLOS 3
The <chn> parameter specifies the channel (1 to 4) to enable. This comand does not
apply to memory channels.
7.4.3.3
ROUTe:OPEN? <chn>
Returns the state of the specified channel.
Example: ROUT:OPEN? 1
Response: 0
The <chn> parameter specifies the channel (1 to 4). If the specified channel is enabled
this query returns 0. If the specified channel is disabled, this query returns 1.
7.4.3.4
ROUTe:OPEN <chn>
Disables the specified channel.
Example: ROUT:OPEN 4
The <chn> parameter specifies the channel (1 to 4) to disable. This command does not
apply to memory channels.
82