Fluke 1529 Technical Manual page 91

Chub-e4 thermometer
Hide thumbs Also See for 1529:
Table of Contents

Advertisement

3
This command only applies to PRT/Thermistor channels. Executing this command on
Thermocouple channels generates a -294, "Incompatible type", error message.
The SENSE suffix, <chn>, specifies the channel (1 to 4), or memory channel (5 to 14).
Specifying the MIN parameter returns a value of 2. Specifying the MAX parameter
returns a value of 4. Specifying the DEF parameter returns a value of 4.
7.4.2.8
SENSe<chn>:RESistance:WIRE <num>|MIN|MAX|DEF
Sets the probe wire setting.
SENS2:RES:WIRE MIN
This command only applies to PRT/Thermistor channels. Executing this command on
Thermocouple channels generates a -294, "Incompatible type", error message.
The SENSE suffix, <chn>, specifies the channel (1 to 4), or memory channel (5 to
14). The <num> parameter must either be 2, 3, or 4. Specifying the MIN, MAX or
DEF parameter sets the wire setting to the minimum, maximum, or default value
respectively.
This command may be password-protected. To access this command:
If the SYST:PASS:CONV? command returns a 0 (probe password-protection is
disabled), the password is not required.
If the SYST:PASS:CONV? command returns a 1 (probe password-
protection is enabled), the password must be successfully entered using the
SYST:PASS:CEN <pass> command (see Section 9.4.11, Password Commands).
7.4.2.9
TRIGger:TIMer? [MIN|MAX|DEF]
Returns the scan sequence timer setting.
TRIG:TIM?
4
Specifying the MIN parameter returns a value of 0.1. Specifying the MAX parameter
returns a value of 3600. Specifying the DEF parameter returns a value of 1. The units
are seconds.
7.4.2.10
TRIGger:TIMer <num>|MIN|MAX|DEF
Sets the scan sequence timer setting.
TRIG:TIM 30
The <num> parameter can be one of the following values: 0.1, 0.2, 0.5, 1, 2, 5, 10, 30,
60, 120, 300, 600, 1800 or 3600. The units are seconds. If a value other than one listed
here is used, the instrument automatically selects the nearest acceptable value that
is less than the value entered. Values greater than 3600 or less than 0.1 are ignored.