1529 Chub-E4 Thermometer Readout
Commands
7.4.8.9
UNIT:TEMPerature?
Returns the system temperature units.
UNIT:TEMP?
C
The response is either C for Celsius, F for Fahrenheit, or K for Kelvin.
7.4.8.10
UNIT:TEMPerature <unit>
Sets the system temperature units.
UNIT:TEMP C
The <unit> parameter is either C or CEL for Celsius, F or FAR for Fahrenheit, or K or
Kelvin for Kelvin. The *RST command sets the units to Celsius.
7.4.9

Communication Interface Commands

The communication interface commands are used for querying and setting serial port
and IEEE configuration parameters.
7.4.9.1
SYSTem:COMMunicate:SERial:BAUD? [MIN|MAX|DEF]
Returns the serial port baud rate setting.
SYST:COMM:SER:BAUD?
2400
Specifying the MIN parameter returns a value of 1200. Specifying the MAX param-
eter returns a baud rate of 19200. Specifying the DEF parameter returns a baud rate of
9600.
7.4.9.2
SYSTem:COMMunicate:SERial:BAUD <baud>|MIN|MAX|DEF
Sets the serial port baud rate setting.
SYST:COMM:SER:BAUD MIN
The <baud> parameter can be one of the following values: 1200, 2400, 4800, 9600,
or 19200. If a value other than one listed is used, the instrument automatically selects
the nearest acceptable value. Values greater than 19200 or less than 1200 are ignored.
Specifying the MIN, MAX or DEF parameter sets the port baud rate setting to the
minimum, maximum, or default value respectively.
7.4.9.3
SYSTem:COMMunicate:SERial:FDUPlex? [MIN|MAX|DEF]
Returns the serial port echo (duplex) setting.
102