Use the LR buttons to select the channel. Press the Enter button to view the cali-
bration parameters for the selected channel. Press the Exit button to cancel and to exit
to the menu.
For PRT/thermistor channels, the parameters C0, C100, C400, C0K, C10K, C100K,
DATE, and DUE are displayed.
The C0, C100, and C400 parameters apply to the PRT probe types and 100W
resistance range. Use these parameters to adjust the measurement at 0, 100 and
400W respectively. These values are in ohms.
The C0K, C10K, and C100K parameters apply to the thermistor probe types and
10KW resistance range. Use these parameters to adjust the measurement at 0K,10K,
and 100KW respectively. These values are in Kohms.
The DATE parameter is the date the current calibration was performed.
The DUE parameter is the date the current calibration expires.
For thermocouple channels, the parameters C0, C100, CRJ, DATE, and DUE are
displayed.
The C0 and C100 parameters adjust the measurement at 0 and 100 mV respectively.
These values are in mV.
The CRJ parameter adjusts the reference junction temperature measurement. This
parameter is in °C.
The DATE parameter is the date the current calibration was performed.
The DUE parameter is the date the current calibration expires.
Press the LR buttons to change a setting and the UD buttons to scroll between
parameters. Press the Enter button to save the new setting. Press the Exit button to
exit to the menu.
6.5.5

System Reset

The SYSTEM RESET function resets all operating parameters to the factory default
values. This function is password-protected (see Section 6.5.3, Password on page 62).
In order to access the system reset, you must enter the correct four-digit password.
When this function is selected, the bottom portion of the display prompts the user to
enter the current password to gain access to this option. (See .Figure 45 on page 63)
By default the password is set to '1529' when the instrument ships from the factory.
We recommend that the user change the default password to protect the password-
protected settings.
Note: Keep your password in a secure location and do not forget it.
To enter the password, press the Enter button. Press the LR buttons to highlight
the digit to change. Use the UD buttons to scroll through the numbers 0-9. Press the
Enter button to save the new setting. Press the Exit button to cancel and to exit to the
menu.
Menu Functions
System Menu
65