1529 Chub-E4 Thermometer Readout
Probe Menu
Figure 15
Use the LR buttons to select the source channel or memory location. Press the
Enter button to save the new setting. Press the Exit button to cancel and to exit to the
menu.
6.2.3

Test Probe

The TEST PROBE function (Figure 16 on this page) allows you to test the probe
characterization algorithm and characterization coefficients for a specific probe. The
probe and probe type must be setup in the EDIT PROBE function prior to entering
the TEST PROBE function. When this function is selected, the bottom portion of the
display shows the probe channel or memory location, the probe serial number, and the
conversion type.
Figure 16
38
24.159C
1
24.142C
3
FROM CH:
PMEMO
P5
SERIAL:
TYPE:
PT100
1529
CHUB E-4 THERMOMETER READOUT
ENTER / MENU
Copy Probe
24.159C
1
24.142C
3
PROBE:
CHAN1
SERIAL:
145-560
CVD
TYPE:
1529
CHUB E-4 THERMOMETER READOUT
ENTER / MENU
Test Probe
24.634C
2
24.015C
4
TO CH:
CHAN1
UNIT
CONTRAST
EXIT
24.634C
2
24.015C
4
UNIT
CONTRAST
EXIT