Logging Commands - Fluke 1529 Technical Manual

Chub-e4 thermometer
Hide thumbs Also See for 1529:
Table of Contents

Advertisement

1529 Chub-E4 Thermometer Readout
Commands
7.4.6.9
DISPlay:WINDow<n>:FIELd<n>:FEED?
Returns the current setting of the specified field for the specified display set.
DISP:WIND1:FIEL2:FEED?
0,16
The WINDow suffix, <n>, specifies the display set (1 to 9). The FIELd suffix, <n>,
specifies the field (1 to 8) within the display set. The response consists of two num-
bers separated by a comma. The first number represents the channel associated with
the field. The second number represents the parameter to be displayed in the specified
field. See Table on page , Field Types, for the field types.
7.4.6.10
DISPlay:WINDow<n>:FIELd<n>:FEED <chn>[,<num>]
Sets the specified field within the specified display set.
DISP:WIND1:FIEL3:FEED 0,10
The WINDow suffix, <n>, specifies the display set (1 to 9). The FIELd suffix, <n>,
specifies the field (1 to 8) within the display set. The <chn> parameter specifies chan-
nel (1 to 4) associated with the field. The <num> parameter specifies the parameter to
be displayed in the specified field. See Table on page , Field Types, for the field types.
7.4.7

Logging Commands

The logging commands are used for logging data automatically and on demand.
7.4.7.1
LOGging:AUTomatic:DELete [<num>|ALL]
Deletes entries from the auto log.
LOG:AUT:DEL 10
Specifying a number in the <num> parameter causes all auto log entries with that label
number (1 to 25) to be deleted. If the ALL parameter is specified or no parameter is
specified, all auto log entries are deleted.
7.4.7.2
LOGging:AUTomatic:COUNt? [MIN|MAX|DEF]
Returns the number of measurements to be stored in the auto log during an auto log
session.
LOG:AUT:COUN?
5000
Specifying the MIN parameter returns a value of 1. Specifying the MAX parameter
value returns a value of 8160. Specifying the DEF parameter returns a value of 8160.
94