ROUT:SCAN:MODE? MIN
0
If 0 is returned, the scan mode is simultaneous. If a 1 is returned the scan mode
is scan. Specifying the MIN parameter returns a value of 0. Specifying the
MAX parameter returns a value of 1. Specifying the DEF parameter returns a
value of 1.
8.4.3.8
ROUTe:SCAN:MODE <num>|MIN|MAX|DEF
Sets the scan mode setting.
ROUT:SCAN:MODE 1
The <num> specifies the scan mode (1 for scan, 0 for simultaneous). Spec-
ifying the MIN, MAX, or DEF parameter sets the mode to the minimum, maxi-
mum, or default, respectively. The *RST command sets scan mode to scan.
8.4.4

Probe Commands

The probe commands are used for querying, setting, and copying the probe
conversion type, conversion parameters, and serial number and for testing the
temperature conversion for a channel. The probe commands for copying and
setting parameters are password protected.
8.4.4.1
CALCulate<chn>:CONVert:CATalog?
Returns a list of conversion types available for the specified channel.
CALC2:CONV:CAT?
"RES","ITS","ITS5","PT","CVD","TRES","TTEM","YSI"
The CALCulate suffix, <chn>, specifies the channel (1 to 4) or memory chan-
nel (5 to 14). Each item in the response list is enclosed in double quotes and
separated from other items by a comma. See Table 7 on page 124, for the avail-
able conversion types.
8.4.4.2
CALCulate<chn>:CONVert:COPY <dest chn>|ALL
Copies the conversion type, sub-ranges (ITS-90), serial number, and character-
ization coefficients from one channel to another.
CALC2:CONV:COPY 8
The CALCulate suffix <chn> specifies the source channel (1 to 4). The <dest
chn> parameter specifies the destination channel (1 to 4). When the copy is
performed, the source channel parameters are not affected. If the destination
channel type is different from the source channel type, the error -294, "Incom-
patible type" is generated. Specifying the ALL parameter copies the source
channel parameters to all compatible channels. The example above copies the
parameters from channel 2 to memory channel PMEM3.