Edit Probe; Figure 14 Edit Probe - Fluke 1529 Technical Manual

Chub-e4 thermometer
Hide thumbs Also See for 1529:
Table of Contents

Advertisement

1529 Chub-E4 Thermometer Readout
Probe Menu
Figure 13
6.2.1

Edit Probe

The EDIT PROBE function allows you to select a probe channel, enter the probe serial
number, and select the conversion type and characterization coefficients and param-
eters. You select the probe channel from one of the four probe channel setups or from
one of the ten memory setups.
Figure 14
The EDIT PROBE function can be password protected to restrict access. See Sec-
tion 6.5.3, Password on page 62, for more information. If this function is password
protected, the user is prompted to enter the password to gain access. If an incorrect
password is entered, the display shows the message PASSWORD INCORRECT and
access is denied. If the correct password is entered or if this function is not password
30
24.159C
1
24.142C
3
PROBE CHARACTERIZATION FUNCTIONS
EDIT
COPY
PROBE
PROBE
1529
CHUB E-4 THERMOMETER READOUT
ENTER / MENU
Probe Menu
24.159C
1
24.142C
3
PROBE:
CHAN-1
SERIAL:
145-560
CVD
TYPE:
1529
CHUB E-4 THERMOMETER READOUT
ENTER / MENU
Edit Probe