Clear Stats; Select Fields; Figure 20 Clear Stats - Fluke 1529 Technical Manual

Chub-e4 thermometer
Hide thumbs Also See for 1529:
Table of Contents

Advertisement

6.3.1

Clear Stats

The CLEAR STATS function clears the statistical registers and resets all statistical cal-
culations for all channels. Refer to Section 6.3.2, Select Fields on page 41, to display
statistical data. When this function is selected, the bottom portion of the display states
that you must press the Enter button to clear the statistical data.
Figure 20
Clear Stats
Press the Enter button to clear the statistical data. Press the Exit button to abort clear-
ing the statistical data and to exit to the menu.
6.3.2

Select Fields

The SELECT FIELDS function (Figure 21 on next page) allows you to select a
display set (1 through 9) to use for displaying information provided by the instrument.
A display set can be customized to display up to eight different fields. See the table of
Field Types below. When this function is selected, the bottom portion of the display
allows you to select the display set and also shows the information that is displayed for
each set. To edit a display set, see Section 6.3.3, Edit Fields on page 43.
24.159C
1
24.142C
3
CLEAR STATISTICAL DATA.
Press ENTER to continue...
1529
CHUB E-4 THERMOMETER READOUT
ENTER / MENU
UNIT
24.634C
2
24.015C
4
CONTRAST
EXIT
Menu Functions
Fields Menu
41

Advertisement

Table of Contents
loading