Print Probe; Default Probe; Figure 17 Print Probe - Fluke 1529 Technical Manual

Chub-e4 thermometer
Hide thumbs Also See for 1529:
Table of Contents

Advertisement

Use the LR buttons to select the channel or memory location. After the channel is
selected, press the Enter button to test the temperature calculation for the probe. You
can enter an input value (W, KW, or millivolts) and the corresponding output value
(e.g. temperature) is shown. Press the Exit button to exit to the menu.
6.2.4

Print Probe

The PRINT PROBE function (Figure 17 on this page) allows you to print probe
characteristics to the serial port. Probe characteristics for a specific channel or memory
location can be printed or for all channel and memory locations. When this function is
selected, the bottom portion of the display shows the probe channel.
Figure 17
Print Probe
Use the LR buttons to select the channel, memory location, or ALL. Use the UD
buttons to scroll between parameters. Press the Enter button to continue and to send
the data to the selected port. A message is displayed stating the channel to print and the
port. Press the Enter button to print the data. Press the EXIT button to cancel and to
exit to the menu.
6.2.5

Default Probe

The DEFAULT PROBE function (Figure 18 on next page) allows you to reset the
probe coefficients for a selected channel, memory location, or ALL to the factory
defaults. When this function is selected, the bottom portion of the display shows the
probe channel.
24.159C
1
24.142C
3
PROBE:
PROBE:
CHAN1
PORT:
SERIAL
1529
CHUB E-4 THERMOMETER READOUT
ENTER / MENU
UNIT
24.634C
2
24.015C
4
CONTRAST
EXIT
Menu Functions
Probe Menu
39

Advertisement

Table of Contents
loading

Table of Contents