7.4.4.4
CALCulate<chn>:CONVert:NAMe <conv>
Sets the conversion type for the specified channel.
CALC2:CONV:NAME CVD
The CALCulate suffix, <chn>, specifies the channel (1 to 4) or memory channel (5
to 14). The <conv> parameter is a mnemonic indicating the conversion type. The ac-
ceptable conversion types depend on the channel type. The conversion types and their
names are listed in Table on page .
This command may be password-protected. To access this command:
If the SYST:PASS:CONV? command returns a 0 (probe password-protection is
disabled), the password is not required.
If the SYST:PASS:CONV? command returns a 1 (probe password-
protection is enabled), the password must be successfully entered using the
SYST:PASS:CEN <pass> command (see Section 9.4.11, Password Commands).
7.4.4.5
CALCulate<chn>:CONVert:PARameter:CATalog?
Returns the names of the probe characterization parameters associated with the current
conversion type for the specified channel.
CALC2:CONV:PAR:CAT?
"RANGE","RTPW","A4","B4","A","B","C","D"
The CALCulate suffix, <chn>, specifies the channel (1 to 4) or memory channel (5 to
14). Each item in the response list is enclosed in double-quotes and separated from
other items by a comma. Each item indicates the name of a characterization parameter
for the current conversion type.If there are no parameters available for the current
conversion type, an empty string is returned. The list of parameters depends on the
selected conversion type. Characterization parameters for the various conversion types
are listed in Table on page , Conversion Types and Parameters.
7.4.4.6
CALCulate<chn>:CONVert:PARameter:VALue? [<param>|ALL]
Returns the value of the specified conversion parameter for the specified channel.
CALC2:CONV:PAR:VAL? RTPW
100.0145
The CALCulate suffix, <chn>, specifies the channel (1 to 4) or memory channel (5 to
14). The <param> parameter is the name of the parameter as given in Table , Conver-
sion Types and Parameters. If the <param> parameter does not apply to the selected
conversion type, an error -221, "Settings Conflict", is generated. Specifying the ALL
parameter returns the name and value of all parameters associated with the specified
channel in a comma delimited list. If no parameter is specified, the ALL parameter is
assumed. The format of the response consists of a parameter name, enclosed in double
Digital Communications Interface