This command is password-protected. To access this command, the password must
be successfully entered using the SYST:PASS:CEN <pass> command (see Section
9.4.11, Password Commands).
7.4.5.6
CALibrate<chn>:PARameter:OFFSet<n>? [MIN|MAX|DEF]
Returns the calibration offset parameter for the specified channel.
CAL1:PAR:OFFS2?
–1.2
The CALibrate suffix, <chn>, specifies the channel (1 to 4). This command does not
apply to memory channels. For PRT/Thermistor channels, the calibration offset suffix
<n> specifies the resistance range parameter for PRTs (1) or Thermistors (2). For Ther-
mocouple channels, the calibration offset suffix <n> is ignored and may be omitted.
Note: For PRT/Thermistor channels, the value returned by this command is always in
ohms.
For PRT/Thermistor channels, specifying the MIN parameter returns a value of -9
when the offset suffix <n> is 1 (for PRTs) and -9000 when the offset suffix <n> is 2
(for Thermistors). For Thermocouple channels, specifying the MIN parameter returns
a value of -9. For PRT/Thermistor channels, specifying the MAX parameter returns a
value of 9 when the offset suffix <n> is 1 (for PRTs) and 9000 when the offset suffix
<n> is 2 (for Thermistors). For Thermocouple channels, specifying the MAX param-
eter returns a value of 9. Specifying the DEF parameter returns a value of 0.
7.4.5.7
CALibrate<chn>:PARameter:OFFSet<n> <num>|MIN|MAX|DEF
Sets the calibration offset parameter for the specified channel.
CAL1:PAR:OFFS2 5
The CALibrate suffix, <chn>, specifies the channel (1 to 4). This command does not
apply to memory channels. For PRT/Thermistor channels, the calibration offset suffix
<n> specifies the resistance range parameter (in ohms) for PRTs (1) or Thermistors
(2). For Thermocouple channels, the calibration offset suffix <n> is ignored and may
be omitted. The <num> paramter specifies the value for the specified offset parameter.
Specifying the MIN, MAX or DEF parameter sets the offset parameter to the mini-
mum, maximum, or default value respectively.
This command is password-protected. To access this command, the password must be
successfully entered using the SYST:PASS:CEN <password> command (see Section
9.4.11, Password Commands).
7.4.5.8
CALibrate<chn>:PARameter:SCALe<n>? [MIN|MAX|DEF]
Returns the calibration scale parameter for the specified channel.
CAL1:PAR:SCAL2?
Digital Communications Interface
Commands
89