8.4.3.6

ROUTe:SCAN?

Returns a comma delimited list of the enabled channels.
ROUT:SCAN?
(@2,3,4)
The list of channels is enclosed in parenthesis and preceded with the @
symbol.
8.4.3.7
ROUTe:SCAN <chn list>
Sets the channels for scanning.
ROUT:SCAN 2,3
The <chn list> parameter is a comma delimited list of the channels (1 to 4) to
enable. Channels that are omitted from this list are disabled. If the <chn list>
parameter is omitted, all channels are disabled. The *RST command sets chan-
nel 1 ON, and all other channels OFF.
8.4.3.8

ROUTe:SCAN:MODE? [MIN|MAX|DEF]

Returns the current scan mode setting.
ROUT:SCAN:MODE? MIN
0
If 0 is returned, the scan mode is simultaneous. If a 1 is returned the scan mode
is scan. Specifying the MIN parameter returns a value of 0. Specifying the
MAX parameter returns a value of 1. Specifying the DEF parameter returns a
value of 1.
8.4.3.9
ROUTe:SCAN:MODE <num>|MIN|MAX|DEF
Sets the scan mode setting.
ROUT:SCAN:MODE 1
The <num> specifies the scan mode (1 for scan, 0 for simultaneous). Spec-
ifying the MIN, MAX, or DEF parameter sets the mode to the minimum, maxi-
mum, or default, respectively. The *RST command sets scan mode to scan.
8.4.4

Probe Commands

The probe commands are used for querying, setting, and copying the probe
conversion type, conversion parameters, and serial number and for testing the
temperature conversion for a channel. The probe commands for copying and
setting parameters are password protected.
8.4.4.1
CALCulate<chn>:CONVert:CATalog?
Returns a list of conversion types available for the specified channel.
8 Digital Communications Interface
Commands
95