Fluke Hart Scientific 1523 User Manual

Thermometer readout
Hide thumbs Also See for Hart Scientific 1523:

Advertisement

Quick Links

1523, 1524
Thermometer Readout
User's Guide
Rev. 891001_EN

Advertisement

Table of Contents
loading

Summary of Contents for Fluke Hart Scientific 1523

  • Page 1 1523, 1524 Thermometer Readout User’s Guide Rev. 891001_EN...
  • Page 2 Fluke Corporation, Hart Scientific Division 799 E. Utah Valley Drive • American Fork, UT 84003-9775 • USA Phone: +1.801.763.1600 • Telefax: +1.801.763.1010 E-mail: support@hartscientific.com www.hartscientific.com Specifications subject to change without notice. • Copyright © 2008 • Printed in USA...
  • Page 3 Limited Warranty & Limitation of Liability Each product from Fluke Corporation, Hart Scientific Division (“Hart”) is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year(s) for the Reference Thermometer. The warranty period begins on the date of the shipment.
  • Page 4 Warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not af- fect the validity or enforceability of any other provision. Fluke Corporation, Hart Scientific Division 799 E. Utah Valley Drive • American Fork, UT 84003-9775 • USA Phone: +1.801.763.1600 •...
  • Page 5: Table Of Contents

    Table of Contents Before You Start ...1 Introduction ... 1 Standard Equipment ... 1 Safety Information ... 2 1.3.1 Warning ...2 1.3.2 Cautions ...3 CE Comments ... 5 1.4.1 EMC Directive ...5 1.4.2 Immunity Testing ...5 Using Clamp-On Ferrites ... 5 Emissions Testing ...
  • Page 6 1523, 1524 Thermometer Readout Figures Figure 1 Clamp-On Ferrite ... 6 Figure 2 Input/Output Connections - 1523 ... 9 Figure 3 Input/Output Connections - 1524 ... 10 Figure 4 Keys ... 11 Figure 5 1523 Menu ... 14 Figure 6 1523 Menu (cont) ... 15 Figure 7 1523 Menu (cont) ...
  • Page 7 Tables Table 1 International Symbols ... 4 Table 2 1523 Input/Output Connections ... 9 Table 3 1524 Input/Output Connections ... 10 Table 4 1523 Key Functions ... 12 Table 5 1524 Key Functions ... 17 Table 6 General Specifications ... 24 Table 7 Millivolt Measurement ...
  • Page 9: Before You Start

    Your Fluke 1523 and 1524 thermometer readout is a handheld, battery operated instrument that measures tem- perature using Platinum resistance Thermometers (PRT), Thermistors, and Thermocouples (TC).
  • Page 10: Safety Information

    A Warning identifies conditions and actions that pose hazard(s) to the user; a Caution identifies conditions and actions that may damage the instrument being used. International symbols used on the reference thermometer and in this manual are explained in Table 1 on page 1.3.1...
  • Page 11: Cautions

    Cautions To avoid possible damage to the reference thermometer or to equipment under test: Do not apply more than the rated voltage, as marked on the reference thermometer, between the inputs, ● or between any input and earth ground (30 V 24 mA max all terminals).
  • Page 12: Table 1 International Symbols

    Safety Information Table 1 International Symbols Symbol Description AC (Alternating Current) AC-DC Battery Complies with European Union directives Double Insulated Electric Shock Fuse Symbol Description PE Ground Hot Surface (Burn Hazard) Read the User’s Guide (Important Information) Canadian Standards Association C-TICK Australian EMC mark The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/EC) mark.
  • Page 13: Ce Comments

    CE Comments 1.4.1 EMC Directive Hart Scientific’s equipment has been tested to meet the European Electromagnetic Compatibility Directive (EMC Directive, 2004/108/EC ). The Declaration of Conformity for your instrument lists the specific stan- dards to which the unit was tested. The instrument was designed specifically as a test and measuring device.
  • Page 14: Emissions Testing

    Emissions Testing The instrument fulfills the limit requirements for Class B. Low Voltage Directive (Safety) In order to comply with the European Low Voltage Directive (2006/95/EC), Fluke equipment has been de- signed to meet the EN 61010-1. probe cable clamp-on ferrite...
  • Page 15: 1.8 Authorized Service Centers

    1.8 Authorized Service Centers Please contact one of the following authorized Service Center to coordinate service on your Fluke product: Fluke Corporation Hart Scientific Division Phone: +1.801.763.1600 Fluke Nederland B.V. Phone: +31-402-675300 Fluke Int'l Corporation - CHINA Phone: +86-10-6-512-3436 Fluke South East Asia Pte Ltd. - SINGAPORE...
  • Page 17: Quick Start

    Quick Start Setup V MAX Figure 2 Input/Output Connections - 1523 CALIBRATION 1523 THERMOMETER READOUT Table 2 1523 Input/Output Connections Name Serial Connector, T1 Power V DC Setup Description Serial interface connector Sensor Connector, Channel 1 External Power adapter connection...
  • Page 18: Figure 3 Input/Output Connections - 1524

    Setup V MAX Figure 3 Input/Output Connections - 1524 CALIBRATION 1524 THERMOMETER READOUT Table 3 1524 Input/Output Connections Name Serial Connector, T1 Connector, T2 Power V DC Description Serial interface connector Sensor Connector, Channel 1 Sensor Connector, Channel 2 External Power adapter connection...
  • Page 19: Figure 4 Keys

    Quick Start Setup Figure 4 Keys...
  • Page 20: Table 4 1523 Key Functions

    Setup Table 4 1523 Key Functions Description Power on or off Yellow Second or Special Function Key Turns the backlight on or off 1st Press: MAX, 2nd Press: MIn, 3rd Press: AVE, 4th Press: STD DEV STATS ° ° Units, °C/°F HOLD 1st press - Holds value on screen "-- HOLD --"...
  • Page 21 Description RECALL 1st Press - Enters Recall menu, 2nd Press - Exits Recall Menu NEXT Moves down to next option on screen. “RESET” - Resets Stats Data STATS ° ° “Ω mV” - Toggles from °C to Ω or Ω to °C (PRT, thermistor), °C to mV or mV to °C (TC) HOLD “TREND”...
  • Page 22: Figure 5 1523 Menu

    1523, 1524 Thermometer Readout Setup Figure 5 1523 Menu...
  • Page 23: Figure 6 1523 Menu (Cont)

    Quick Start Setup Figure 6 1523 Menu (cont)
  • Page 24: Figure 7 1523 Menu (Cont)

    1523, 1524 Thermometer Readout Setup Figure 7 1523 Menu (cont)
  • Page 25: Table 5 1524 Key Functions

    Table 5 1524 Key Functions Description Power on or off Yellow Second or Special Function Key Turns the backlight on or off 1st Press: Max, 2nd Press: Min, 3rd Press: Ave, 4th Press: STD DEV STATS ° ° Units, °C/°F HOLD 1st press - Holds value on screen "-- HOLD --"...
  • Page 26 Setup Description Selects highlighted selection, Saves a new selection. ENTER RECALL 1st press - Enters Recall Menu, 2nd press - Exits Recall Menu NEXT Moves down to next option on screen. “RESET” - Resets Stats Data STATS ° ° “Ω mV” - Toggles from °C to Ω or Ω to °C (PRT, thermistor), °C to mV or mV to °C (TC) HOLD “TREND”...
  • Page 27: Figure 8 1524 Menu

    Quick Start Setup Figure 8 1524 Menu...
  • Page 28: Figure 9 1524 Menu (Cont)

    1523, 1524 Thermometer Readout Setup Figure 9 1524 Menu (cont)
  • Page 29: Figure 10 1524 Menu (Cont)

    Quick Start Setup Figure 10 1524 Menu (cont)
  • Page 30: Figure 11 1524 Menu (Cont)

    1523, 1524 Thermometer Readout Setup Figure 11 1524 Menu (cont)
  • Page 31: Figure 12 1524 Menu (Cont)

    Quick Start Setup Figure 12 1524 Menu (cont)
  • Page 32: Specifications

    Specifications Figure 13 1524 Menu (cont) Specifications Specifications are based on a one year calibration cycle and apply from 13 °C to 33 °C unless stated otherwise. All specifications assume a five minute warm up period. Table 6 General Specifications Operating Temperature †...
  • Page 33: Table 7 Millivolt Measurement

    3 AA alkaline batteries Power requirements 12 V dc universal power supply Size 96 x 200 x 47 mm (3.75 x 7.9 x 1.86 inches) Weight 0.65 kg (1.4 lb) Safety EN 61010-1:2001, CAN/CSA C22.2 No. 61010.1-04 Environmental conditions for all specifications: 13 °C to 33 °C †...
  • Page 34: Table 9 Ohms Measurement, Thermistor

    Specifications Table 9 Ohms Measurement, Thermistor Ohms Range 200 Ω to 50 kΩ 50 kΩ to 500 kΩ Temperature Coefficient ( –10 °C to 13 °C , +33 °C to 60 °C): 0.002 %/°C + 0.1 Ω (0 Ω to 50 kΩ) 0.06 %/°C + 0.1 Ω...
  • Page 35 Measure Type Range Accuracies (ITS-90) –200 °C to 0 °C 0.36 °C 0 °C to 900 °C 0.23 °C –20 °C to 0 °C 0.26 °C 0 °C to 400 °C 0.25 °C 400 °C to 1400 °C 0.22 °C –200 °C to 0 °C 0.72 °C 0 °C to 1300 °C...
  • Page 36: Table 11 Temperature, Rtd Ranges, And Accuracies (Its-90)

    Specifications Table 11 Temperature, RTD Ranges, and Accuracies (ITS-90) Accuracy ± °C 4 Wire Probe ± 0.011 at –100 °C ± 0.015 at 0 °C ± 0.019 at 100 °C ± 0.023 at 200 °C ± 0.031 at 400 °C ±...