Display Commands - Fluke 1529 Technical Manual

Chub-e4 thermometer
Hide thumbs Also See for 1529:
Table of Contents

Advertisement

1529 Chub-E4 Thermometer Readout
Commands
The RJC suffix, <chn>, specifies the channel (1 to 4). This command does not apply
to memory channels. The <num> parameter specifies the RJC value for the specified
channel. Specifying the MIN, MAX, or DEF parameter sets the RJC parameter to the
minimum, maximum, or default value respectively.
This command is password-protected. To access this command, the password must
be successfully entered using the SYST:PASS:CEN <pass> command (see Section
9.4.11, Password Commands).
7.4.6

Display Commands

The display commands are used for querying and setting the display resolution, lamp
brightness, decimal format, window setting, and window fields.
7.4.6.1
DISPlay:RESolution? [MIN|MAX|DEF]
Returns the display resolution setting.
DISP:RES?
2
Specifying the MIN pamameter returns a value of 0. Specifying the MAX parameter
returns a value of 4. Specifying the DEF parameter returns AUT.
7.4.6.2
DISPlay:RESolution <num>|AUT|MIN|MAX|DEF
Sets the display resolution setting.
DISP:RES 3
The <num> parameter specifies the number of decimal places to display. This pa-
rameter can be 0, 1, 2, 3, or 4. Specifying the MIN, MAX or DEF parameter sets the
display resolution to the minimum, maximum, or default value respectively. Specify-
ing the AUT parameter sets the display resolution to automatic.
7.4.6.3
DISPlay:LAMP? [MIN|MAX|DEF]
Returns the lamp brightness setting.
DISP:LAMP?
0
Specifying the MIN parameter returns a value of 0. Specifying the MAX parameter
returns a value of 3. Specifying the DEF parameter returns a value of 0.
7.4.6.4
DISPlay:LAMP <bool>|<num>|MIN|MAX|DEF
Sets the lamp brightness setting.
92