Status Commands - Fluke 1529 Technical Manual

Chub-e4 thermometer
Hide thumbs Also See for 1529:
Table of Contents

Advertisement

SYST:PASS:NEW 1212
The <pass> parameter must be a 4-digit number (0000 to 9999). The password is re-
quired to access the indicated settings. This command is password protected. To access
this command, the password must be successfully entered using the SYST:PASS:CEN
<pass> command (see Section 9.4.11, Password Commands).
7.4.12

Status Commands

The status commands are used for querying and setting the state of the instrument
registers. For details on the operation of the status registers refer to the IEEE-488.2
and SCPI standards.
7.4.12.1
*CLS
Clears the status registers.
*CLS
7.4.12.2
*ESE?
Returns the Event Status Enable Register.
*ESE?
255
7.4.12.3
*ESE <num>|MIN|MAX|DEF
Sets the Event Status Enable Register.
*ESE 48
7.4.12.4
*ESR?
Returns the Event Status Register.
*ESR?
0
7.4.12.5
*SRE?
Returns the Service Request Enable Register.
*SRE?
0
Digital Communications Interface
Commands
109