Fluke 1529 Technical Manual page 108

Chub-e4 thermometer
Hide thumbs Also See for 1529:
Table of Contents

Advertisement

1529 Chub-E4 Thermometer Readout
Commands
7.4.7.15
LOGging:DEMand:FREE?
Returns the number of demand log entries free and stored.
LOG:DEM:FREE?
87,12
The response consists of two numbers separated by a comma. The first number indi-
cates the number of demand log entries still available. The second number indicates
the number of entries already stored.
Note: Every time a demand log reading is stored, the first entry in the log for that
stored reading is the data label and the date. All subsequent entries in the demand log
consist of the channel, measurement value, units, and time.
7.4.7.16
LOGging:DEMand:LABel? [MIN|MAX|DEF]
Returns the current demand log label setting.
LOG:DEM:LAB?
6
Specifying the MIN parameter returns a value of 1. Specifying the MAX parameter
returns a value of 25. Specifying the DEF parameter returns a value of 1.
7.4.7.17
LOGging:DEMand:LABel <num>|MIN|MAX|DEF
Sets the current demand log label.
LOG:DEM:LAB 3
The <num> parameter specifies the demand log label number (1 to 25) to select. Spec-
ifying the MIN, MAX, or DEF parameter sets the demand log label to the minimum,
maximum, or default respectively.
7.4.7.18
LOGging:DEMand:POINt?
Returns the number of demand log entries already stored.
LOG:DEM:POIN?
21
Specifying the MAX parameter returns the total number of entries that can be stored in
the demand log.
7.4.7.19
LOGging:DEMand:PRINt [<num>|ALL [,<port>]]
Prints the demand log entries.
98

Advertisement

Table of Contents
loading

Table of Contents