ment period to 2 seconds. This command may be password protected (see Sec-
tion 8.2.10, Password Commands).
8.2.5

Channel Commands

The channel commands are used for switching channels on and off.
8.2.5.1
ROUTe:CLOSe? <chn>
Returns the state of the specified channel.
Example: ROUT:CLOS? 1
Response: 0
The <chn> parameter specifies the channel (1 or 2). A value of 1 is returned if
the specified channel is on. A value of 0 is returned if the specified channel is
off.
8.2.5.2
ROUTe:CLOSe <chn>
Switches on the specified channel.
Example: ROUT:CLOS 1
The <chn> parameter specifies the channel (1 or 2) to switch on. This com-
mand may be password protected (see Section 8.2.10, Password Commands)
8.2.5.3
ROUTe:OPEN? <chn>
Returns the state of the specified channel.
Example: ROUT:OPEN? 1
Response: 0
The <chn> parameter specifies the channel (1 or 2). A value of 1 is returned if
the specified channel is off. A value of 0 is returned if the specified channel is
on.
8.2.5.4
ROUTe:OPEN <chn>
Switches off the specified channel.
Example: ROUT:OPEN 1
The <chn> parameter specifies the channel (1 or 2) to switch off. This com-
mand may be password protected (see Section 8.2.10, Password Commands)
8.2.6

Calibration Commands

The calibration commands are used to access the sensor calibration parameters.
The calibration commands are always password protected.
8 Digital Communications Interface
Commands