1551A Ex/1552A Ex SCPI Commands Instruction Sheet Introduction This Instruction Sheet explains the 1551A Ex/1552A Ex (Product) remote SCPI commands. Remote Communications RS-232 Port Operation The RS-232 configuration is fixed as follows: • 8 data bits • 1 stop bit •...
Page 2
SCPI Commands Remote Communications Table 1. List of All Commands Command Command Explanation Returns the instrument identification string that indicates the *IDN? manufacturer, model number, serial number, and code firmware version. Sets the last calibration date for the device. The <year> parameter is a CAL:DEV:DATE four-digit number, 2000 to 2099.
Page 3
SCPI Commands Remote Communications Command Reaction Returns the calculated temperature in Celsius for the given sensor CALC:CONV:TEST? <res> reading, or “0.0,OL” if the result is out of range. The parameter <res> is the sensor reading in ohms. Returns the last measurement in units according to the UNIT:TEMP FETC? setting.