Trigger[:Sequence]:Delay <Numeric_Value; Trigger[:Sequence]:Delay; Trigger[:Sequence]:Timer <Numeric_Value; Trigger[:Sequence]:Timer - Fluke 1560 Black Stack User Manual

Thermometer readout
Table of Contents

Advertisement

6.5.2.11
TRIGger[:SEQuence]:DELay <numeric_value>
This command sets the minimum time period between measurements. It is
equivalent to setting the DELAY value with the MEAS soft-key function in the
INPUT menu. The <numeric_value> parameter value has the range 0 to 32767.
The command also acceps the words MIN (0), MAX (32767), and DEF (0) for
the <numeric_value> parameter. The *RST command sets this value to 0.
6.5.2.12

TRIGger[:SEQuence]:DELay?

This query command returns the measurement DELAY period.
6.5.2.13
TRIGger[:SEQuence]:TIMer <numeric_value>
This command sets the scan sequence timer. It is equivalent to setting the SEQ
TIMER parameter with the MEAS soft-key Function in the INPUT menu. The
<numeric_value> has the range 0 to 10000. This command also accepts the
words MIN (0), MAX (10000), and DEF (0) for the <numeric_value>.
6.5.2.14

TRIGger[:SEQuence]:TIMer?

This query command returns the value of the scan sequence timer. The words
MIN, MAX, or DEF can also be appended to the command to read the corre-
sponding limits and default.
6.5.3

Input Channel Commands

Example command: TRIG:DEL 10
Example command: TRIG:DEL?
Example response: 0